61-4375-54 DC Signal Source SS7012
toner

61-4375-54 DC Signal Source SS7012

model: 61-4375-54
brand: HIOKI E. E. CORPORATION
category: toner

Product Description

Product Model: 61-4375-54Product Brand: HIOKI E. E. COR […]

Contact Us

Hotline

18866622123

WeChat

WeChat ID: 18866622123

WeChat QR Code

Product Details

Product Model: 61-4375-54
Product Brand: HIOKI E. E. CORPORATION

Features

  • DC signal generator capable of simultaneous generation and measurement
  • Improved stability and reduced calibration cost (compared to our company 7011 conventional machines)
  • Loop test of instrumentation system (4-20 mA)
  • Checking temperature control equipment and distributor
  • Thermoelectric power generation (TC) 8 types
  • Maintenance of production facilities as calibrator and testing of electronic equipment
  • Use as electronic load (maximum DC25mA)

Spec

  • Size: 104W x 180H x 58Dmm
  • Weight: 710g (including nickel hydrogen rechargeable battery Z0101)
  • Basic specification: Accuracy warranty period of 1 year, Post-adjustment accuracy warranty period of 1 year
  • Output system: Bipolar sink and source method
  • Constant Voltage (CV) 2.5V: 0 - ±2.5000V (±0.03% of setting ±300μV, Resolution 100μV), 25V: 0 - ±25.000V (±0.03% of setting ±3mV, Resolution 1mV)
  • Constant current (CC): 25mA: 0 - ±25.000mA (±0.03% of setting ±3μA, Resolution 1μA)
  • Thermoelectromotive force (TC: 0℃) K: -174.0 - 1372.0℃ (±0.05% of setting ±0.5℃, Resolution 0.1℃), Generation range and accuracy are determined individually for other E, J, R, S, N
  • Thermoelectromotive force (TC: RJ) K: -174.0 - 1372.0℃ (±0.05% of setting ±1.0℃, Resolution 0.1℃), Generation range and accuracy are determined individually for other E, J, R, S, N
  • Standard resistance (Rs): 100Ω (±0.2Ω)
  • Memory generation: Maximum number of steps 20, Generation time interval: 1 - 99s (CV, CC, TC mode)
  • Measurement function:
  • Voltage: 2.5V: 0 - ±2.8000V (±0.03%rdg. ±300μV, Resolution 100μV, Input resistance 1MΩ), 25V: 0 - ±28.000V (±0.03%rdg. ±3mV, Resolution 1mV, Input resistance 1MΩ)
  • Current: 25mA: 0 - ±28.000mA (±0.03%rdg. ±3μA, Resolution 1μA, Input resistance 30Ω or lower)
  • Temperature: -25.0 - 80.0℃ (±0.5℃ at 23 ±5℃, Resolution 0.1℃, When using 9184)
  • Sampling rate: Approximately 1.67 times/s
  • Additional function: Zero adjustment, Overload display, USB communication, Monitor
  • Power supply: AC adapter 9445-02 (rating 100 - 240VAC, 9VA), Nickel hydrogen rechargeable battery Z0101 (HR6 x 4, 2500mAh, 6VA), Continuous *** time 170 minutes or more, or N alkali dry cell (LR6 x 4, 6VA)
  • Accessories: Input cord 9168 x 1, Test lead L9170-10 x 1, F*** x 1, AA alkaline dry cell (LR6) x 4, Instruction manual x 1
  • *We recomm*** to *** in combination with AC adapter or charging type battery and charger.
  • SS7012

Related product recommendations

You may also be interested in the following products