![1-5366-01-22 [Discontinued]UV-visible Spectrophotometer with Pre-shipment Inspection Certificate SefiIUV-1240](https://www.jgauge.com/wp-content/uploads/2025/08/3c4297b82666eded94bda4d1c6d237f1.webp)
Analyzer, Testing Equipment
1-5366-01-22 [Discontinued]UV-visible Spectrophotometer with Pre-shipment Inspection Certificate SefiIUV-1240
Product Description
Product Model: 1-5366-01-22Product Brand: AS ONE Corpor […]
Contact Us
Product Details
Product Model: 1-5366-01-22
Product Brand: AS ONE Corporation
Features
- In addition to photometrics, functions for concentration measurement and quantitative analysis (factor method, one-point calibration curve method, and multi-check calibration curve method) are standard.
- Spectrum measurement is provided as standard, and it is easy to check the measurement wavelength (absorption peak automatic detection function).
- Measurement data can be stored on a dedicated IC card and managed on a PC using dedicated software (sold separately).
- Advanced spectrum and quantitative measurements are possible.
Spec
- Body Sefi IUV -1240
- *A measuring cell is not included.
- Measuring wavelength range: 190.0 ~ 1100.0 nm
- Spectral bandwidth: 5 nm
- Wavelength display: 0.1 nm units
- Wavelength setting: 0.1 nm unit (1 nm unit when wavelength scanning area is set)
- Wavelength accuracy: +/- 0.1 nm
- Wavelength repeatability: +/- 0.3 nm
- Light source switching: selectable from the following 3 1: Automatic switching of light source in conjunction with wavelength Wavelength: 295 ~ 364 nm (selectable in 1 nm increments) Recomm***ed wavelength: 340 nm 2: Use only halogen lamp (W1) without switching 3: Use only deuterium lamp (D2) without switching
- Metering: Single-beam metering
- Photometric range: Absorbance/-0.3 to + 3.0 Abs, Transmittance/0.0 ~ 200%
- Photometric accuracy: +/- 0.005 Abs at 1.0 Abs using NIST 930 D filter +/- 0.003 Abs at 0.5 Abs
- Sample chamber dimensions: 110 x 230 x 105 mm (partially 155 mm high), for cell with optical path length of 10 mm
- Power supply: 100 ~ 120 VAC 50/60 Hz 160 VA
- External dimensions/weight: 416 x 379 x 274 mm/11 kg
- Inspection contents App***ix: Inspection inspection documents
- Confirmation of baseline, wavelength accuracy and transmittance of 50%
- App***ix: Inspection Certificate documents
- *Products with Inspection Certificate are Azwan's original service that includes inspection of control and electric engine personnel and issuance of inspection documents at the time of shipment.
Related product recommendations
You may also be interested in the following products
61-0250-83 Opticap Gamma Compatible XL10 Durapore 0.22μm GC 3/4In TC/TC 1/Pk 1ST KVGLG10FF1
model: 61-0250-83
brand: Merck Millipore Corporation
61-0386-25 [Discontinued]Fmoc-Lys(Fmoc)-OH 852041 25G 8520410025
model: 61-0386-25
brand: Merck Millipore Corporation
61-0226-18 Polygard-CR Cart Filt 30In. 100μm Code 7 Silicone 6/Pk 6 x 1ST CR9973006
model: 61-0226-18
brand: Merck Millipore Corporation