Features
- For indep***ent quadruple calorimetry and process optimization experiments.
- Heat of reaction is measured with high precision and sensitivity by heat flow and cascade control. The actual heat of reaction can be measured from low to high temperatures by mass balancing.
- The four reaction vessels are indep***ently temperature-controlled (P.I.D control), and the four cooling channels allow adjustment of heating and cooling times. It is also equipped with a low temperature tank to cope with abnormal temperature rise.
Spec
- Impeller stirring
- Reaction vessel/Number of bottles: Separable flasco/1 ~ 4 bottles
- Synthetic scale: 50 mL/100 mL/250 mL (choice)
- Temperature control range (aluminum block temperature): - 20 ~ 200 C (4 x individual temperature control) * A cooling water circulator is required.
- Temperature control accuracy: +/- 0.5 K (- 20 ~ 100 deg. C)/+/- 0.1 K (100 ~ 200 deg. C)
- Speed range: Impeller/0, 125 ~ 1000 rpm (4 x individual stirring), Stirrer/0, 100 ~ 1000 rpm (4 x individual stirring)
- Maximum heating rate: Tr < 5.3 K/min, Tj < 19.3K/min(5°C→95°C)
- Maximum cooling rate: Tr < 7.2 K/min, Tj < 27.8K/min(95°C→5°C)
- Maximum heat removal: 0.74 ~ 1.1 W/g (200 mL water in a glass reactor)
- Temperature control: Reactor, Jacket, Insulation, Gradient, Isothermal, Heating and Cooling P.I.D, Cascade
- Dropping Control: Weight Dropping or Deposition Dropping (1 ~ 4 reams)
- Stirring control: Impeller stirrer (selective)
- pH control (optional): 0 ~ 14 pH, acid or base (1-liquid control or 2-liquid control)
- Calorimetry: Heat flow, specific heat, reaction heat calculation, calibration heater, with CalGraph for recalculation and analysis
- Setting and recording: Setting and recording for each experiment, automatic recording of parameters such as temperature and dripping
- Tr*** graph export: Real-time pen, with function to retrieve historical pen display and saved data to Excel
- Safety features: Emergency cooling function/Automatic operation at reactor control End-step setting, earth leakage/overcurrent breaker with emergency program setting, aluminum block indep***ent overrise detector, motor overload protection, software self-diagnosis function (Upper temperature, temperature rise speed, motor overload), communication error
- Operating ambient temperature range: 5 ~ 35 degrees C
- External dimensions: 642 x 510 x 380 mm
- Weight: approximately 53 kg
- Power input/voltage: 20 A, 2.0 kVA AC 100 V 50/60 Hz
- *A personal computer is included.
- *Temperature and pH sensors are sold separately.
- *The temperature unit K is Kelvin (absolute temperature).
- Code: 254030
Related product recommendations
You may also be interested in the following products
62-8412-12 Casein Kinase I Inhibitor III,PF-4800567 218719-5MG
model: 62-8412-12
brand: CBC(Merck)
62-8519-18 Anti-CRTC2, clone 5B10 MABN695
model: 62-8519-18
brand: Sigma Aldrich Japan(Millipore)
62-8351-93 Opticap Gamma Compatible XLT10 Express SHR 0.1μm 1-1/2in. TC-5/8in. HB 1/pk KVEPG1TTH1
model: 62-8351-93
brand: Sigma Aldrich Japan(Millipore)