63-1395-46 Thin film evaporator MF-10A
laboratoryinstruments

63-1395-46 Thin film evaporator MF-10A

model: 63-1395-46
brand: TOKYO RIKAKIKAI CO., LTD.

Product Description

Product Model: 63-1395-46Product Brand: TOKYO RIKAKIKAI […]

Contact Us

Hotline

18866622123

WeChat

WeChat ID: 18866622123

WeChat QR Code

Product Details

Product Model: 63-1395-46
Product Brand: TOKYO RIKAKIKAI CO., LTD.

Features

  • Tabletop Type
  • Suitable for concentration of foaming and heat-sensitive samples
  • Suitable for concentration of heat-denaturable and foaming samples.

Spec

  • Speed range: 140 ~ 1200 rpm (continuously variable)
  • Evaporative capacity: 1140 mL/h (water) when hot water circulates at 70 degrees C and vacuum is 4.0 kPa (30 torr)
  • Reached vacuum: vacuum pump capacity + 13.3 hPa (1 Torr)
  • Viscosity ***d: about 1000 cP
  • Temperature setting range: room temperature + 5 ~ 90 degrees C
  • Rotation setting and display: Volume setting and digital display
  • Timer function: Control the sample injection volume by setting the ON/OFF time of the solenoid valve.
  • Motor: DC Brushless motor output 60 W
  • Rotating shaft seal: Vacuum seal
  • Evaporation tube: Jacket-type evaporation area: 0.04 m2
  • Cooler: Vertical double serpentine cooling area: 0.3 m2
  • Receiving flask: 1L concentrate side, 1L evaporator side, with drain cock
  • Heating source: External hot water circulator (sold separately)
  • Wetted part material: Hard glass, Teflon (R), SUS 304
  • Connection diameter: outside diameter of nozzle for cooling/hot water circulation: 12 mm, outside diameter of suction/sample nozzle: 8 mm
  • Sample injection system: On-Off type solenoid valve with needle type and timer set
  • Operating ambient temperature range: 5 ~ 35 degrees C
  • External dimensions: 625 x 400 x 1416 mm
  • Weight: 40 kg
  • Power input/power voltage: 2 A, 200 VA/AC 100 V 50/60 Hz A separate power supply is required for the heating source.
  • *Vacuum pumps, cooling water circulators, etc. are required for ***.
  • *Evaporation capacity varies dep***ing on conditions such as decompression conditions, hot water circulation tank, and cooling water temperature.
  • *The degree of vacuum during operation dep***s on the cooling water temperature, the decompression device and the sample.
  • Code: 117440
  • *Shipping, carrying-in, installation and assembly costs are required separately. Please inquire.

Related product recommendations

You may also be interested in the following products

1-3096-02 Thermo-Hygrostat (Temperature Setting: From 10℃) LH-114
laboratoryinstruments
2-1948-11 Highly Pure Water Generation Unit PR0G0T0S2 Proguard TS2 Filter Cartridge PROG0T0S2
laboratoryinstruments
1-5449-11 Solution Temperature Controller ETS-D5
laboratoryinstruments