laboratoryinstruments
63-1239-54 Shaking culture apparatus, small size TVS062CA
Product Description
Product Model: 63-1239-54Product Brand: ADVANTEC Featur […]
Contact Us
Product Details
Product Model: 63-1239-54
Product Brand: ADVANTEC
Features
- The growth of microorganisms was automatically measured under a constant temperature environment.
- The absorbance and temperature of 6 samples can be measured automatically by shaking the culture at room temperature +5°C to 70°C.
- Up to 6 devices can be connected and 36 samples can be measured.
Spec
- Operation form: constant temperature
- Operating temperature range: R.T. + 5 °C, to 70 °C
- Temperature Adjustment Accuracy: ± 0.2 °C
- Shaking angle: 60 °
- Shaking speed: 20 rpm to 70 rpm (1 rpm unit)
- Setting items and setting ranges
- Preparation time: 0 or 5 minutes to 300 minutes
- Shaking time: 1 min. to 99 min. (ON time), 0 min. to 99 min. (OFF time)
- Absorbance determination interval: 1 minute to: 99 minutes
- Operation time: 1 hour to 249 hours 59 minutes
- Ambient operating temperature: 5 °C to 35 °C (non-condensing)
- Absorbance measurement range: O.D. value not more than 2.0
- temperature controller: Microcomputer PID Control
- Temperature Sensor: Platinum Temperature Measuring Resistor Pt 100 Ω
- Heater: Silicone rubber heater
- Heater capacity: 100 W
- Shaking Method: Monode Type Arc Reciprocating Motion
- Culture block part: Aluminum conductor
- Culture Block Culture Tube Insertion Part: Small L-type Culture Tube 6 pieces (Or test tube, vial bottle)
- Light source of detector: tungsten lamp 6 pieces
- Detector Light Receiver: Silicon Photodiode 6 pieces
- Detector filter: Glass filter 660 nm, 6 pieces
- record
- Data is recorded and stored in separately sold personal computer and printers.
- Up to 6 devices can be connected.
- Data processing is also possible.
- Safety device: Self-diagnostic function (Overheating, temperature sensor disconnection, heater disconnection), indep***ent overheat protection device, overcurrent breaker
- External input/output signal terminal: RS -232 C (Output of temperature and absorbance data by personal computer)
- external dimensions (Installation Space): W 410 mm x D 325 mm x H 160 mm (W 410 mm x D 406 mm x H 160 mm)
- power supply (50Hz/60Hz): AC100V 3A
- Mass: about 10 kg
- Accessories: Small L-type Culture Tube 6 pieces, Culture Tube Cap 6 pieces, power cord 1 piece