3-3936-11 [Discontinued]Fume Hood Standard (with FRP fan with inverter) 1200 x 830 x 2250mm ABSF-1200INV
gasmonitors

3-3936-11 [Discontinued]Fume Hood Standard (with FRP fan with inverter) 1200 x 830 x 2250mm ABSF-1200INV

model: 3-3936-11
brand: AS ONE Corporation
category: gasmonitors

Product Description

Product Model: 3-3936-11Product Brand: AS ONE Corporati […]

Contact Us

Hotline

18866622123

WeChat

WeChat ID: 18866622123

WeChat QR Code

Product Details

Product Model: 3-3936-11
Product Brand: AS ONE Corporation

Features

  • Since an inverter is mounted on the exhaust fan side, it is very convenient to be able to flexibly change the exhaust air volume and static pressure even after installation.
  • The air quantity display function is equipped as standard, and it is possible to check the exhaust condition and work.

Spec

  • Model number: ABSF -1200 INV
  • Specifications: Standard
  • external dimension (mm): 1200 x 830 x 2250
  • inner size (mm): 950 x 605 x 1200
  • Weight (About kg): 305
  • Exterior: Steel (powder coating finish)
  • Interior: chemical resistant Heat-Resistant Board (Flame-Retardant material/phenolic laminated sheet or glass fiber mixed polyester)
  • Work Plane: Epoxy Plastic top board Integral Part (32 mm Atsushi)
  • Lifting sash: tempered glass (5 mm Atsushi) with balance weight opening restriction key (400 mm Height), Handle/ABS Plastic
  • Hydrant: chemical coating Remote Operated Water Connection G 1/2 (Stop Female thread)
  • drain port: Plastic Drainage connector 40 A socket with a built-in sink and trap
  • Outlet: AC100V -15 A2 grounded
  • lighting: Hf fluorescent lamps
  • Operating Switch: Touch Panel Lighting/Fan
  • Safety Function: Exhaust Air air volume Display
  • Exhaust Connections: PVC (Plastic vinyl chloride ) with Volume Damper Phase Flanged
  • power supply connection part: Terminal block with electric box
  • Exhaust fan: Upper-mounted FRP motor with inverter
  • *Shipping, carrying-in, installation, assembly, etc. are required separately. Please contact us.
  • *If the duct distance is long and the number of turns is large, the fan ability may be insufficient.
  • exhaust
  • Work Sash Opening Height (mm): 400
  • Controlled wind speed (m/s): 0.5
  • air volume (m3/min): 12
  • Duct Connector: 200 A
  • Required duct diameter (Φ mm): 200
  • exhaust fan
  • Type: 100 KS5-O
  • power supply: 3 Φ 200 V
  • Motor (kW): 0.4
  • ancillary equipment
  • Plumbing (place): 1
  • fluorescent lamps (bottle): 2 x 16 W
  • Outlet (place): 2