toner
67-4915-94 Mg6000Series Wall-Mounted Suction Type Oxygen Meter (Transmission: 0~10 Mv, Acidity Meter Mode I, Display: 0.0~25.0%O2 Fixed, Transmission Output: Supports 0.0~25.0%O2) MG6001-000
Product Description
Product Model: 67-4915-94Product Brand: CHINO CORPORATI […]
Contact Us
Product Details
Product Model: 67-4915-94
Product Brand: CHINO CORPORATION
Features
- It is a suction type oxygen meter to measure the oxygen concentration in the furnace.
- MG6000 is an oximeter with a built-in suction pump.
- Measurement gas at high or low temperature in places where oxygen meters or oxygen sensors cannot be directly attached can be sucked by a pipe and measured at around normal temperature.
- Low concentration measurement gas can be sucked to measure oxygen concentration in a furnace or equipment at a distance.
- Measurement of hot or cold gases
- Since a suction pump and a flowmeter are built in, measuring gas can be radiated in the middle of pipes and piping and connected to an oxygen meter at around normal temperature.
- Since a filter containing activated carbon is built in, trace amounts of toxic gas and dust can be removed by the filter.
- An optional filter unit containing activated carbon can be added dep***ing on the operating environment.
- high precision
- It has a high accuracy of ±0.5%O2, and the calibration gas connection port is provided in the body to facilitate calibration with reference gas and maintain accuracy.
- Excellent stability and long life
- It ***s a high-performance compact zirconia solid electrolyte oxygen sensor that has a long life and is easy to maintain, so the accuracy is stable for a long time.
- Small and thin for easy operation
- Both the filter unit and the measuring unit are small and thin and easy to install.
- zirconia oximeter
- Stabilized zirconia is known as an oxygen ion conductor.
- The stabilized zirconia is ***d as a substrate, and a platinum electrode and an oxygen diffusion limiting hole are formed on one surface of the substrate, and a platinum heater is formed on the other surface.
- In order to obtain practical oxygen ion conductivity, the substrate is heated to several hundred °C by energizing the heater. When a voltage is applied between the electrodes, a current with oxygen ions as carriers flows through the substrate.
- Due to the oxygen diffusion limiting hole formed on the cathode, the amount of oxygen flowing into the cathode is limited, and a flat region (limit current) is observed in the current-voltage characteristics.
- Since the limiting current varies in proportion to the oxygen concentration, it is possible to determine the oxygen concentration in the atmosphere by measuring the limiting current.
Spec
- Transmission signal output: 0~10 mV (output resistance approx. 10 Ω)
- alarm setting
- Oxygen deficiency meter mode Type I, 18% lower limit, 26% upper limit
- Buzzer, Lamp, Relay Output With Output Holding
- Display range: 0.0~25.0%O2 fixed
- Transmission output range: output corresponding to 0.0~25.0%O2
- Measurement Principle: Zirconia Solid Electrolyte Method
- Air sampling method: suction pump type
- Response time: 90% Response within 1 minute
- Measurement value display: Digital display with 3-digit yellow-green LED
- Measuring range (resolution): 0.0~25.0 (0.1)%O2
- Accuracy rating: +/- 0.5%O2 +/- 1digit in a dry gas of oxygen and nitrogen
- Transmission method: output to 0.0~25.0 measurement indications
- Alarm function: Measured value alarm ...
- Alarm Setting Code: Anoxic Meter Mode
- Lower warning value: 18% fixed
- Upper warning value: fixed at 26%
- Relay power yes/no: yes
- Buzzer with/without: Yes
- Lamp lit/not lit: yes
- Output holding yes/no: yes
- Detection element abnormality alarm: When the heater current to the oxygen detection element is abnormal, relay output is ON, buzzer is blown, and lamp is lit regardless of the content of the measured value alarm.
- Power failure warning: When the power supply to the tester is OFF or when the f*** is broken, relay output is ON.
- Alarm relay output: number of output circuits (1), output capacity (Resistive Load 110 V AC 0.2 A, 24 V DC 1 A/Inductive Load 110 V AC 0.1 A, 24 V DC 0.5 A)
- Warm-up time: 5 minutes
- Note)Warning signal is OFF and transmission signal is not output during warm-up.
- Operating Temperature Range: 0~40°C
- Operating humidity range: 80%rh or less
- Power: 100 VAC 50/60Hz
- Power Consumption: Approx. 13VA
- Mass: Approx. 4.6 kg
- Mounting Method: Body Bolt Mounting
- ISO 14001 Certified JQA-EM2414
- ISO 9001 Certification JQA-0656
- External Dimensions (Excluding Protrusions)
Related product recommendations
You may also be interested in the following products
65-4293-89-20 Sealed Pressure Gauge (60Φ) Embedded D Frame (Mounting Hole) with 0.1 MPa with Inspection Certificate BA15-271
model: 65-4293-89-20
brand: NAGANO KEIKI Co., Ltd.
65-3628-43 Diaphragm Type Pressure Gauge Screw Type Diaphragm SUS316L Encapsulated Liquid Silicon A Type φ75 76φ 0~1.6 PK-1 AU75 G1/2B 1.6MPA S25C SUS304
model: 65-3628-43
brand: Daiichi Keiki Mfg
65-3592-29 HNT General Purpose Pressure Gauge for Steam/M A Type φ100 G3/8 0~0.06MPA AMT38-1000.06MPA
model: 65-3592-29
brand: Daiichi Keiki Mfg