laboratoryinstruments
68-5095-48 Cooling water circulation device CA-1331
Product Description
Product Model: 68-5095-48Product Brand: TOKYO RIKAKIKAI […]
Contact Us
Product Details
Product Model: 68-5095-48
Product Brand: TOKYO RIKAKIKAI CO., LTD.
Spec
- Cooling/Circulation System: Air-Cooling/Sealed Circulation System
- Performance
- Temperature Setting Range: -20°C~20°C (Without Heater)
- Temperature adjustment accuracy: ±2°C
- Cooling Capacity (at Liquid Temperature): 10°C: 1190 W (1020kcal/h), 0°C: 880 W (758kcal/h), -10°C: 630 W (542kcal)
- External Circulation Capacity (50 Hz/60Hz): Maximum Flow 12/14Lmin, Maximum Pump Head 9.5/13 L (Discharge Pressure: 0.09/0.13 MPa)
- Function
- Temperature control: refrigerator ON/OFF control
- Temperature Setting/Display: Sheet Key Input/Digital Display, Minimum Digit 1°C
- Safety Functions: Leakage/Overcurrent Breaker, Temperature Controller Self-Diagnosis Function, Chiller Overload Relay, High Pressure Switch, Chiller Protection Timer, Circulation Pump Thermal Protector
- Accessory: Ball Valve
- External Input/Output: Communication Terminal for Vacuum Controller
- Configuration
- Air Cooled Output 775 W/R290
- Cooling Coil: Copper-Nickel Plated
- Standard
- External Circulation Nozzle: One-touch Connector Inner Diameter 10 mm, One-touch Hose Nozzle Outer Diameter 10 mm
- Water Tank Material: Stainless Steel
- Water Tank Dimensions (mm), Capacity: 270 W x 270D x 175 H, 12.8 L (Liquid Amount 10 L)
- Operating ambient temperature range: 5~35°C
- External Dimensions (mm), Weight: 460 W x 430 (500) D x 570 (598) H, About 43 kg
- Power Input/Power Voltage: 11 A, 1.1kVA/100 V 50/60Hz
- *Performance is measured at room temperature of 20°C, clean water, rated supply voltage of 50 Hz, and no load.
- *The cooling capacity is ±10% of the display capacity.
- *The external circulation capacity is ±10% of the display capacity.
- *The cooling temperature reached at room temperature of 35°C is -10°C.
- *Use a low-temperature heating medium when setting the liquid temperature to 10°C or lower.
- *Dimensions in () include projections.
- *Pure water cannot be circulated.
- *Do not block both sides of the product during installation. If one side is blocked, the cooling capacity decreases by about 10%.
Related product recommendations
You may also be interested in the following products
65-2060-25 Fluororesin (PTFE) Sphere Value Type 2.38Φ 100 Pieces Pack NR0346-001
model: 65-2060-25
brand: Flon Chemical